A medical fridge only protects its contents if you can trust the temperature it shows. A thermometer or data logger that reads two degrees out can hide a fridge that has been running warm for weeks, and that can mean discarded stock, recalled patients and an awkward accreditation finding. Medical fridge calibration checks that the readings are right, that the fridge holds its range and that you have records to prove it.
Which fridges need calibration?
- Vaccine fridges in GP practices, pharmacies, councils and aged care. These must hold +2°C to +8°C under the National Vaccine Storage Guidelines (Strive for 5). See our vaccine fridge calibration service.
- Medication fridges holding insulin, biologics and other temperature sensitive medicines on wards and in clinics.
- Specimen and pathology fridges used to hold samples before transport.
- Blood fridges in hospitals and blood banks, which have their own requirements under AS 3864 and local blood management policies.
Domestic fridges are not designed for this work. Purpose built medical or pharmaceutical fridges hold temperature far more evenly, and most accreditation bodies expect them for vaccine and medicine storage.
What medical fridge calibration involves
- Checking the fridge’s own display, thermometer or data logger against a calibrated reference probe, so you know the reading you rely on every day is accurate.
- Checking the temperature across the storage space, because shelves near the door, the back wall or the cooling plate can run warmer or colder than the display suggests.
- Testing alarms, so the fridge warns you when it drifts out of range.
- Electrical safety testing of the fridge as part of your AS/NZS 3551 equipment program.
- A report you can file for your accreditation assessor or immunisation provider audit.
How often should medical fridges be calibrated?
Once a year is a common baseline for calibration of the fridge’s thermometer or data logger, alongside the daily minimum and maximum checks your staff record. Calibrate again after a repair, a relocation or any cold chain breach, and follow any shorter interval set by the fridge or logger manufacturer, your state health department or your accreditation body. Our cold chain compliance guide covers vaccine fridge requirements in more detail.
Signs your fridge needs checking now
- The display and a second thermometer disagree.
- Minimum or maximum readings regularly sit close to the edge of the range.
- Ice is building up, or the door seal is damaged.
- The alarm has not sounded in a long time, or sounds without reason.
- The fridge has been moved, repaired or had a power outage.
On site medical fridge calibration
Nova Biomedical calibrates medical fridges on site for GP practices, hospitals, pharmacies and aged care providers across Queensland, New South Wales, Western Australia and New Zealand. We can check every fridge in the same visit as your medical equipment testing, so one booking covers your whole site. We are ISO 9001:2015 certified.
Call 1300 723 900 or request a quote.
Frequently asked questions
Is a data logger enough on its own?
A data logger records what happened, but only if it reads accurately. It still needs calibration against a reference, and someone still needs to review its records and act on alarms.
Do you need to empty the fridge?
Usually not. Calibration is done with the fridge in normal use, so your stock stays in the cold chain.
What temperature should a vaccine fridge be?
Between +2°C and +8°C, aiming for +5°C, under the National Vaccine Storage Guidelines.